Advanced Concepts of Orbital Reconstruction

A Unique Attempt to Scientifically Evaluate Individual Techniques in Reconstruction of Large Orbital Defects
Open AccessPublished:December 18, 2020DOI:https://doi.org/10.1016/j.cxom.2020.10.003

      Keywords

      Key points

      • The use of a cadaver model allows assessment of the effect of individual steps of computer-assisted surgery on orbital implant positioning.
      • The accuracy and consistency of implant positioning without computer-assisted surgery are low, which enphasizes the need for intraoperative control.
      • 3D virtual surgical planning leads to a better implant position through enhanced insight in anatomy and fracture details, and detailed knowledge on the ideal implant position.
      • Deviations from the ideal implant position can be identified with intraoperative imaging and the implant’s position can be improved accordingly.
      • Real-time navigation proves to be the safest, fastest and most accurate method of intraoperative navigation for orbital reconstruction.

      Introduction

      In most, if not all, clinical research papers and presentations on novel techniques in orbital reconstruction, combinations of different techniques have been used to highlight technological progress and increased quality of surgical outcomes in terms of predictability and reliability.
      • Gellrich N.-C.
      • Schramm A.
      • Hammer B.
      • et al.
      Computer-assisted secondary reconstruction of unilateral posttraumatic orbital deformity.
      • Bly R.A.
      • Chang S.-H.
      • Cudejkova M.
      • et al.
      Computer-guided orbital reconstruction to improve outcomes.
      • Rana M.
      • Chui C.H.K.
      • Wagner M.
      • et al.
      Increasing the accuracy of orbital reconstruction with selective laser-melted patient-specific implants combined with intraoperative navigation.
      • Zimmerer R.M.
      • Ellis E.
      • Aniceto G.S.
      • et al.
      A prospective multicenter study to compare the precision of posttraumatic internal orbital reconstruction with standard preformed and individualized orbital implants.
      • Chen C.-T.
      • Pan C.-H.
      • Chen C.-H.
      • et al.
      Clinical outcomes for minimally invasive primary and secondary orbital reconstruction using an advanced synergistic combination of navigation and endoscopy.
      • De Cuyper B.
      • Abeloos J.
      • Swennen G.
      • et al.
      Intraoperative navigation and cone beam computed tomography for restoring orbital dimensions: a single-center experience.
      In the Amsterdam University Medical Centers, The Netherlands, a comprehensive research started in 2014 with the development of a human cadaver model for assessment of orbital reconstruction. This model was used in sequence to evaluate every single technological step in the process of orbital reconstruction. This article is a resume of this series of investigations and publications in which the additive value of individual steps used in computer-assisted surgery, including various intraoperative navigation techniques, in orbital reconstruction has been assessed.
      The interventions evaluated were as follows:

      Materials and methods

      Materials and methods of assessment were almost identical in all studies. Fixed human cadaver heads were obtained from the Department of Medical Biology, Section of Clinical Anatomy and Embryology of the Amsterdam University Medical Centers. Computed tomography (CT) scans of the cadaver heads were performed at baseline (with intact orbits, T0) (SOMATOM Sensation 64; Siemens Healthineers, Erlangen, Germany). Scan parameters included collimation of 20 × 0.6 mm, 120 kV, 350 mAs, pitch 0.85, field of view 30 cm, matrix 512 × 512, reconstruction slice thickness of 0.75 mm with overlapping increments of 0.4 mm, bone kernel H70s, and bone window W1600 L400. The orbital floor and medial wall were fully exposed through a standard transconjunctival incision and retroseptal preparation. Following the Jaquiéry classification, complex orbital defects (class III–IV) were created with piezoelectric surgery (Mectron SpA, Carasco, Italy).
      • Jaquiéry C.
      • Aeppli C.
      • Cornelius P.
      • et al.
      Reconstruction of orbital wall defects: critical review of 72 patients.
      After creation of the orbital defects (T1), and postoperatively after implant placement (T2), consecutive CT scans were acquired using the aforementioned scanning protocol. After T2 imaging was acquired, the orbital reconstruction plates were removed, and the screw holes were covered with dental filling material (DuraLay; Reliance Dental Mfg Co, Worth, IL, USA) to make them invisible. In all studies, preformed orbital titanium mesh implants (KLS Martin, Tuttlingen, Germany) were used; the stereolithographic models (STL files) of these implants were imported in the iPlan Cranial software environment (version 3.0.5; Brainlab AG, Munich, Germany). The optimal position of the implant was virtually planned and served as the target position during reconstruction. The faces of the human cadaver heads were covered to prevent recognition in consecutive studies, and the various studies were carried out with time intervals of several months to years.
      Validation studies were performed to investigate the accuracy and precision of different methods for orbital reconstruction. The acquired implant position was compared with the target implant position of the planning. For this purpose, a reference frame for the preformed orbital implant was created (Fig. 1).
      • Schreurs R.
      • Dubois L.
      • Becking A.G.
      • et al.
      Quantitative assessment of orbital implant position-a proof of concept.
      This reference frame made quantification of the difference between acquired and planned implant position possible: the individual rotational (roll, pitch, and yaw, in degrees) and translational (in millimeters) deviations served as the outcome parameters for all studies. In studies 1 to 5, 19 orbits were reconstructed by 2 surgeons; 10 orbits were reconstructed by 1 surgeon in study 6, and 20 orbits (of another cadaver group) were reconstructed by 2 surgeons in study 7. The results of studies 2 to 5 were compared with the baseline results (study 1). Studies 6 and 7 were compared with the results of study 5, because image-guided navigation could be considered the reference method for evaluating these enhanced navigation methods.
      Figure thumbnail gr1
      Fig. 1Orbital implant positioning frame. The roll (red), pitch (green), and yaw (blue) rotation around the reference axes are visualized.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD and Leander Dubois, MD, DMD, PhD.)

      Conventional transconjunctival reconstruction (study 1)

      In the first surgical session, all orbits were reconstructed without further technological additions, with a transconjunctival approach. Preformed orbital implants were positioned according to the best judgment of the surgeon and fixed with osteosynthesis screws. It is important to know that the surgeon had no access to information on the preoperative virtual planning, but only to a multiplanar visualization and 3-dimensional (3D) reconstruction of the CT scan acquired after creation of the orbital defects (T1). The information available to the surgeon and hardware used are visualized in Fig. 2. This study acted as a baseline study for comparison of implementation of the technologies evaluated.
      Figure thumbnail gr2
      Fig. 2Information available to the surgeon during conventional transconjunctival reconstruction. The multiplanar reconstruction and 3D reconstruction of the T1 CT were accessible; no additional hardware was used.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with transsinusoidal endoscopy (study 2)

      The orbital reconstructions in this surgical session were performed with the additional use of a transsinusoidal endoscope. The standard transconjunctival approach was again used to position the implant. To facilitate inspection with an endoscope, a gingivobuccal incision and a 5-mm antrostomy were created in the canine fossa concavity using a Piezotome (Mectron SpA). The sinus mucosa was removed, so the defect and position of the implant could be visualized from the maxillary sinus with a 30°endoscope (KARL STORZ SE & Co KG, Tuttlingen, Germany). In Fig. 3, the available information during surgery and the hardware used are visualized. The results of this session were compared with the baseline study (study 1).
      Figure thumbnail gr3
      Fig. 3Additional information at hand in reconstruction with transsinusoidal endoscopy (left), and hardware used (right). The multiplanar reconstruction and 3D reconstruction from study 1 (see ) were accessible as well.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with virtual surgical planning (study 3)

      In the third session, the surgeons had access to a 3D virtual computer environment and to virtual planning tools in iPlan Cranial, such as segmentation and mirroring. Information on the ideal position of the orbital implant, which had been planned with the help of a skilled and experienced technical physician, was available during reconstruction (Fig. 4). The surgeons were challenged to put the implant in the best position according to the virtual plan. The results of reconstruction with preoperative 3D virtual planning were compared with the baseline results (study 1).
      Figure thumbnail gr4
      Fig. 4The virtual surgical planning for reconstruction that was visible to the surgeon during reconstruction in study 3. No hardware tools were used in the workflow.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with intraoperative imaging (study 4)

      An intraoperative CT scan was made after reconstruction of the orbits in this study. A scanning protocol identical to the T0, T1, and T2 studies was used for intraoperative imaging. The implant position was verified by the surgeon on the information seen in Fig. 5 and, if required, the orbital implant position was corrected. The procedure and the scans were repeated until the surgeon was satisfied with the resulting implant position. The results were compared with baseline (study 1).
      Figure thumbnail gr5
      Fig. 5Intraoperative scanning feedback. The result of intraoperative imaging is visualized on the screen (the necessary hardware is also shown). Based on these images, the surgeon could decide whether adjustment of the implant position was necessary.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with intraoperative image-guided navigation (study 5)

      All orbits were reconstructed with assistance of intraoperative image-guided navigation (Kolibri; Brainlab AG). The preoperative virtual surgical plan was linked to the intraoperative position of the cadaver head (registration).
      • Eggers G.
      • Mühling J.
      • Marmulla R.
      Image-to-patient registration techniques in head surgery.
      • Metzger M.C.
      • Hohlweg-Majert B.
      • Schön R.
      • et al.
      Verification of clinical precision after computer-aided reconstruction in craniomaxillofacial surgery.
      • Widmann G.
      • Stoffner R.
      • Bale R.
      Errors and error management in image-guided craniomaxillofacial surgery.
      • Schramm A.
      • Suarez-Cunqueiro M.M.
      • Rücker M.
      • et al.
      Computer-assisted therapy in orbital and mid-facial reconstructions.
      • Yu H.
      • Shen S.G.
      • Wang X.
      • et al.
      The indication and application of computer-assisted navigation in oral and maxillofacial surgery—Shanghai’s experience based on 104 cases.
      The cadaver head was tracked using a skull reference marker array (Brainlab AG). After a stable position of the orbital implant was achieved intraoperatively, the navigation pointer was moved along the implant. The result was verified on the multiplanar views: the surgeon could compare the trajectory of the pointer with the contour of the planned position of the implant, which was shown as an overlay (Fig. 6). Results were compared with the baseline study (study 1).
      Figure thumbnail gr6
      Fig. 6Image-guided navigation, where information is collected through the use of a navigation pointer. The pointer is moved along the contour of the implant; the position of the pointer is shown in the multiplanar views, where it can be compared with the planned contour of the implant.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with intraoperative marker-based navigation (study 6)

      In this study, intraoperative navigation was again used, but feedback was obtained by indicating orientation markers in the implant design. Three markers were embedded in the preformed implant design, in a triangular fashion (Fig. 7).
      • Dubois L.
      • Essig H.
      • Schreurs R.
      • et al.
      Predictability in orbital reconstruction. A human cadaver study, part III: implant-oriented navigation for optimized reconstruction.
      ,
      • Gander T.
      • Essig H.
      • Metzler P.
      • et al.
      Patient specific implants (PSI) in reconstruction of orbital floor and wall fractures.
      The markers were indicated as landmarks in the virtual surgical planning, which allowed comparison between the current position of the marker (indicated with the navigation pointer) and the planned position. This feedback was both visual, to obtain information about the direction of displacement, and quantitative, on the amount of displacement (as seen in Fig. 8). The implant-guided navigation workflow results were compared with the results of the image-guided navigation workflow (study 5).
      Figure thumbnail gr7
      Fig. 7Navigation markers embedded in the preformed implant (arrowheads).
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)
      Figure thumbnail gr8
      Fig. 8Feedback obtained if the pointer is positioned at one of the marker points. In the multiplanar views, the position of the marker in relation to the landmark in the virtual surgical planning is compared for guidance. Quantitative feedback on the deviation between pointer position and planned marker position is also provided.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with intraoperative real-time navigation (study 7)

      For this study, a real-time navigation workflow was developed and used. An instrument (Titanium Orbital Positioner) was designed that attached to the preformed orbital implant and was compatible with tracking by the navigation system. The cohesion between implant, instrument and navigation tracking allowed calculation of the implant position from the instrument’s position, even during positioning of the implant. Real-time feedback on the current implant position in comparison to the planned position was provided, both visually and quantitatively, and the surgeon could be steered to the correct implant position under navigation guidance. The instrument attachment to the preformed implant and the feedback obtained during positioning are shown in Fig. 9. The reconstruction was compared with reconstruction with implant-oriented navigation (study 6).
      Figure thumbnail gr9
      Fig. 9Real-time navigation feedback and hardware. The designed instrument attached to the implant (and could be released after fixation). This made visualization in 3D possible (with the color of the implant changing according to the amount of deviation still present) as well as quantitative feedback on all positioning parameters.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Results

      The results are all presented in later discussion in the same manner; statistical workflows, however, differed in some situations because of study design or scientific insights over the years. The outcome of the Orbital Implant Positioning Frame (shown in Fig. 1) parameters (roll, pitch, yaw, and translation) are provided in tables, and graphic representation of the mean obtained implant position with each technique (in relation to the planned position) is provided in the figures.

      Conventional transconjunctival reconstruction (study 1)

      The rotational and translational deviations of the acquired implant position in relation to the planned implant position are provided in Table 1. The mean and standard deviations of the positioning parameters show that there is minimal control in positioning of the orbital implant: the mean rotation around the z-axis (yaw) is larger than 15° and the mean translation is 5 mm. A maximum value of 47.6° is seen for the yaw. Precision, the consistency of positioning, is also low, as seen by the large standard deviations. Visual appraisal of the average acquired position of the implant (Fig. 10) illustrates the size of the quantitative deviations found. These results underline the need for intraoperative control during reconstruction.
      Table 1Rotational and translational deviation in conventional reconstruction (baseline)
      Conventional
      MeanSD
      Roll (deg)−9.89.1
      Pitch (deg)−1.33.1
      Yaw (deg)17.810.9
      Translation (mm)5.02.2
      The mean and standard deviations (SD) of the rotations (roll, pitch, yaw) and translation are given.
      Abbreviation: deg, degrees.
      Figure thumbnail gr10
      Fig. 10Angulated, top, and frontal view of the mean acquired position of the orbital implant with conventional reconstruction (orange). The planned implant position is visualized in gray. A deviation in yaw can be best appraised in the top view, whereas roll is best seen in the frontal view.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with transsinusoidal endoscopy (study 2)

      The rationale behind endoscopic-assisted orbital reconstruction is that a bidirectional view is created for the surgeon: the implant’s position may be visually assessed both transconjuctivally and, with the aid of an endoscope, transsinusoidally.
      • Dubois L.
      • Jansen J.
      • Schreurs R.
      • et al.
      Predictability in orbital reconstruction: a human cadaver study. Part I: endoscopic-assisted orbital reconstruction.
      The resulting implant positions are quantified in Table 2 and visualized in Fig. 11. The results suggest that there is no significant improvement with the use of an endoscope compared with the baseline. Although there are some unrelated advantages regarding exhibition of the procedure for training surgeons,
      • Bevans S.E.
      • Moe K.S.
      Advances in the reconstruction of orbital fractures.
      ,
      • Moe K.S.
      • Murr A.H.
      • Wester S.T.
      Orbital fractures.
      accuracy and consistency of implant placement are again low.
      Table 2Quantification of endoscopic-assisted implant positioning, compared with conventional implant positioning
      ConventionalEndoscopeP Value
      MeanSDMeanSD
      Roll (deg)−9.89.1−7.15.6.19
      Pitch (deg)−1.33.1−0.33.7.24
      Yaw (deg)17.810.914.511.8.14
      Translation (mm)5.02.24.22.1.19
      Figure thumbnail gr11
      Fig. 11Mean acquired position with endoscopic-assisted orbital reconstruction (blue) compared to the planned position (grey).
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with virtual surgical planning (study 3)

      With virtual surgical planning available, the surgeon is able to check the information in the planning during the reconstruction of the orbit. In Table 3, the results are compared with the baseline study: roll, yaw, and translation are significantly more accurate with 3D virtual surgical planning. The standard deviations are smaller than in the baseline reconstructions as well, demonstrating improved precision. The addition of 3D virtual planning is considered to be superior to conventional reconstruction with multi-planar reconstruction information from a CT scan. Additional anatomic landmarks of the preoperative 3D planning, enhanced insight in the extent of the fracture, and the detailed knowledge of the ideal implant position help the surgeon to achieve a better reconstruction result (in terms of implant position).
      • Gellrich N.-C.
      • Schramm A.
      • Hammer B.
      • et al.
      Computer-assisted secondary reconstruction of unilateral posttraumatic orbital deformity.
      ,
      • Schmelzeisen R.
      • Gellrich N.C.
      • Schoen R.
      • et al.
      Navigation-aided reconstruction of medial orbital wall and floor contour in cranio-maxillofacial reconstruction.
      ,
      • Mahoney N.R.
      • Peng M.Y.
      • Merbs S.L.
      • et al.
      Virtual fitting, selection, and cutting of preformed anatomic orbital implants.
      The mean acquired implant position is visualized in Fig. 12.
      Table 3Results of reconstruction with virtual planning versus results of conventional planning
      ConventionalVirtual PlanningP Value
      MeanSDMeanSD
      Roll (deg)−9.89.1−4.53.5.01
      Pitch (deg)−1.33.1−1.12.2.78
      Yaw (deg)17.810.95.26.8<.01
      Translation (mm)5.02.22.61.7<.01
      Figure thumbnail gr12
      Fig. 12Average position of the implant for reconstruction with preoperative virtual planning (green). The planned position is visualized in gray. Compared with and , the average obtained position resembles the planned position more closely.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with intraoperative imaging (study 4)

      With intraoperative imaging, major deviations from the ideal position can be identified and may be altered intraoperatively, to avoid a second procedure and potential complications.
      • Jaquiéry C.
      • Aeppli C.
      • Cornelius P.
      • et al.
      Reconstruction of orbital wall defects: critical review of 72 patients.
      ,
      • Stuck B.A.
      • Hülse R.
      • Barth T.J.
      Intraoperative cone beam computed tomography in the management of facial fractures.
      ,
      • Wilde F.
      • Schramm A.
      Intraoperative imaging in orbital and midface reconstruction.
      The position of the orbital implant improves significantly in terms of roll, yaw, and translation compared with the baseline study (Table 4), showing the advantageous effect of an intraoperative imaging during reconstruction. In Fig. 13, the improvement on implant positioning yielded by intraoperative imaging is shown. Significant improvement in yaw, the parameter with the largest deviation, is also recorded if the position after the first reconstructive attempt and the final CT scan are compared (Table 5), demonstrating the learning effect of intraoperative imaging and the guidance it provides.
      Table 4Quantification of orbital implant position obtained with intraoperative imaging
      ConventionalIntraoperative ImagingP Value
      MeanSDMeanSD
      Roll (deg)−9.89.1−2.54.2.01
      Pitch (deg)−1.33.1−1.83.7.37
      Yaw (deg)17.810.96.87.8<.01
      Translation (mm)5.02.23.01.4<.01
      Compared with conventional reconstruction, significant improvements for all parameters but pitch are seen.
      Figure thumbnail gr13
      Fig. 13Obtained implant position after reconstruction with intraoperative imaging (yellow). The mean of the final implant positions is visualized.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)
      Table 5Comparison of first result and final result in the cadaver group with adjustments after the first intraoperative image
      First Intraoperative CTFinal Intraoperative CTP Value
      MeanSDMeanSD
      Roll (deg)−2.18.2−1.53.6.78
      Pitch (deg)−1.22.9−1.94.5.48
      Yaw (deg)10.213.25.610.1.02
      Translation (mm)3.61.42.91.0.16

      Reconstruction with intraoperative image-guided navigation (study 5)

      Image-guided navigation feedback may be generated immediately after the implant is first positioned. From comparison of the pointer’s trajectory with the implant’s contour in the multiplanar views, the surgeon is able to evaluate the need for adjustment of the position of the implant.
      • Gellrich N.-C.
      • Schramm A.
      • Hammer B.
      • et al.
      Computer-assisted secondary reconstruction of unilateral posttraumatic orbital deformity.
      ,
      • Bly R.A.
      • Chang S.-H.
      • Cudejkova M.
      • et al.
      Computer-guided orbital reconstruction to improve outcomes.
      ,
      • Schramm A.
      • Suarez-Cunqueiro M.M.
      • Rücker M.
      • et al.
      Computer-assisted therapy in orbital and mid-facial reconstructions.
      After adjustments are made, their effect is again assessed directly afterward. This short feedback loop for control and improvement of implant position yields significant improvement in acquired implant position in comparison to the baseline reconstruction (Table 6). Roll, yaw, and translation are further improved through image-guided navigation. Fig. 14 demonstrates the mean position with image-guided navigation and the planned position of the implant.
      Table 6The effect of image-guided navigation on translation and rotation of the resulting implant position
      ConventionalImage-Guided NavigationP Value
      MeanSDMeanSD
      Roll (deg)−9.89.1−2.34.8<.01
      Pitch (deg)−1.33.1−1.12.2.77
      Yaw (deg)17.810.98.88.1<.01
      Translation (mm)5.02.23.31.6<.01
      The baseline results, conventional reconstruction, are used for comparison.
      Figure thumbnail gr14
      Fig. 14Mean position of the implant after image-guided navigation reconstruction (red). The planned implant position is also visualized (gray).
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)

      Reconstruction with intraoperative marker-based navigation (study 6)

      Although image-guided navigation provides direct feedback after positioning and leads to better outcome in terms of positioning, the feedback itself could be improved: moving of the probe along the implant means that the surgeon needs to evaluate the implant’s position through interpretation of continuously shifting multiplanar views.
      • Schreurs R.
      • Dubois L.
      • Becking A.G.
      • et al.
      Implant-oriented navigation in orbital reconstruction. Part 1: technique and accuracy study.
      The thought behind marker-based navigation is that more static, and quantitative, feedback is obtained from the marker positions; the triangular fashion makes interpretation of the 3D position feasible. Marker-based navigation is compared with image-guided navigation (study 5) and proves to be significantly more accurate in positioning an orbital implant in translation and yaw (Fig. 15, Table 7). The precision is improved as well, as seen from the smaller standard deviations. With marker-based navigation, intraoperative handling thus becomes more intuitive.
      Figure thumbnail gr15
      Fig. 15Marker-based navigation result (purple). The mean position resembles the planned position (gray) closely.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)
      Table 7The effect of image-guided navigation on translation and rotation of the resulting implant position
      Image-Guided NavigationMarker-Based NavigationP Value
      MeanSDMeanSD
      Roll (deg)−2.34.8−2.33.1.16
      Pitch (deg)−1.12.2−2.22.8.52
      Yaw (deg)8.88.16.08.1.02
      Translation (mm)3.31.61.40.7<.01
      The baseline results, conventional reconstruction, are used for comparison.

      Reconstruction with intraoperative real-time navigation (study 7)

      The real-time navigation workflow leads to very accurate positioning of the orbital implant, with small deviations compared with the planned position (Fig. 16, Table 8). Compared with the marker-based approach, positioning improves in terms of translation, roll, and yaw. This improvement may be explained by the real-time nature of the feedback.
      • Schreurs R.
      • Dubois L.
      • Ho J.
      • et al.
      Implant-oriented navigation in orbital reconstruction part II: preclinical cadaver study.
      In image-guided or marker-based navigation, the surgeon can either obtain feedback or adjust the implant’s position and needs to switch between these processes. In real-time navigation, adjustment of the implant position is possible under navigation guidance. Feedback is presented in a true 3D fashion. Next to an improvement in implant position, surgery time is reduced and appraisal of the feedback by the surgeon is higher.
      Figure thumbnail gr16
      Fig. 16Mean position of the implant obtained with the real-time navigation workflow. The result (pink) is virtually indistinguishable from the planned position (gray), indicating excellent accuracy.
      (Courtesy of Ruud Schreurs, MSc, Alfred G. Becking, MD, DMD, PhD, FEBOMS, Jesper Jansen, MD, DMD, PhD, and Leander Dubois, MD, DMD, PhD.)
      Table 8Linear mixed model estimates for the effect of technique on the outcome parameters
      Marker-Based NavigationReal-Time NavigationP Value
      Model EstimateModel Estimate
      Roll (deg)3.22.0.03
      Pitch (deg)1.51.3.53
      Yaw (deg)3.42.2.01
      Translation (mm)1.81.3<.01
      The P values are generated from Likelihood Ratio tests between a model with and without navigation.

      Discussion

      These studies have been designed and executed to assess the relevance of individual parts of modern technologic advancements for orbital reconstruction, instead of the grouped results of computer-assisted surgery (3D virtual planning, intraoperative imaging, and navigation). Significant progress in reliability and predictability has been achieved by introducing 3D virtual preoperative planning in which surgeons are thought to gain better insight in local anatomy, defect size, and complexity. This enhanced insight in turn generates helpful additional anatomic landmarks, leading to improvement in implant positioning.
      • Gellrich N.-C.
      • Schramm A.
      • Hammer B.
      • et al.
      Computer-assisted secondary reconstruction of unilateral posttraumatic orbital deformity.
      ,
      • Schmelzeisen R.
      • Gellrich N.C.
      • Schoen R.
      • et al.
      Navigation-aided reconstruction of medial orbital wall and floor contour in cranio-maxillofacial reconstruction.
      ,
      • Mahoney N.R.
      • Peng M.Y.
      • Merbs S.L.
      • et al.
      Virtual fitting, selection, and cutting of preformed anatomic orbital implants.
      The same is true for traditional intraoperative imaging with CT, which also demonstrates a significant improvement in implant positioning, and it is hoped, decreases the need for secondary procedures.
      • Stuck B.A.
      • Hülse R.
      • Barth T.J.
      Intraoperative cone beam computed tomography in the management of facial fractures.
      ,
      • Wilde F.
      • Schramm A.
      Intraoperative imaging in orbital and midface reconstruction.
      Three methods of intraoperative navigation have been assessed in combination with 3D virtual planning. Image-guided navigation significantly improves implant positioning compared with traditional reconstruction; however, marker-based navigation increases implant positioning even further and brings the mean translation within the calibration error. Using the Advanced Concepts of Orbital Reconstruction research protocol, the more user-friendly real-time navigation proves to be the safest, fastest, and most accurate way to place an orbital implant into the ideal position. The instrument for orbital implant positioning has not been clinically introduced yet: it is a proof of concept and in the process of further development.
      The goal to assess different techniques for discriminative outcomes of individual steps in orbital reconstruction seems to be achieved, even with the restrictions in the materials and methods: small series, few surgeons, and no clinical circumstances (human cadaver models). Studies 1 to 5 were designed to be one-on-one comparisons of the assessed techniques, but studies 6 and 7 had a different design. Care has been taken to make the outcomes comparable, for instance, through repeating the marker-based reconstruction for the cadaver group of study 7. Still, interpretation of these results and translation to a clinical setting warrant caution.
      The focus of the studies in this article, and of many other studies, is on orbital implant positioning.
      • Zimmerer R.M.
      • Ellis E.
      • Aniceto G.S.
      • et al.
      A prospective multicenter study to compare the precision of posttraumatic internal orbital reconstruction with standard preformed and individualized orbital implants.
      ,
      • Metzger M.C.
      • Hohlweg-Majert B.
      • Schön R.
      • et al.
      Verification of clinical precision after computer-aided reconstruction in craniomaxillofacial surgery.
      ,
      • Stuck B.A.
      • Hülse R.
      • Barth T.J.
      Intraoperative cone beam computed tomography in the management of facial fractures.
      ,
      • Essig H.
      • Dressel L.
      • Rana M.
      • et al.
      Precision of posttraumatic primary orbital reconstruction using individually bent titanium mesh with and without navigation: a retrospective study.
      ,
      • Rana M.
      • Holtmann H.
      • Kanatas A.N.
      • et al.
      Primary orbital reconstruction with selective laser melted core patient-specific implants: overview of 100 patients.
      Outcome evaluation and presentation lead one to believe that implant positioning will be improved tremendously by using these novel technologies. Of course, the overarching goal for orbital reconstruction is not to improve orbital implant position, but to improve the quality of life for patients with orbital wall fractures. Surgery itself is an additional trauma to the orbit and its contents and should be well indicated and balanced with the anticipated outcomes of nonsurgical or less-invasive treatment options.
      • Zimmerer R.M.
      • Gellrich N.-C.
      • von Bülow S.
      • et al.
      Is there more to the clinical outcome in posttraumatic reconstruction of the inferior and medial orbital walls than accuracy of implant placement and implant surface contouring? A prospective multicenter study to identify predictors of clinical outcome.
      ,
      • Jansen J.
      • Dubois L.
      • Maal T.J.J.
      • et al.
      A nonsurgical approach with repeated orthoptic evaluation is justified for most blow-out fractures.
      Surgery and reconstruction with orbital implants have a distinct place in treatment of orbital wall fractures. When indicated, orbital implant positioning should be performed as accurately and precisely as possible, to improve bulb position, ocular movements, orbital volumes, and esthetic outcome.
      • Jaquiéry C.
      • Aeppli C.
      • Cornelius P.
      • et al.
      Reconstruction of orbital wall defects: critical review of 72 patients.
      ,
      • Schmelzeisen R.
      • Gellrich N.C.
      • Schoen R.
      • et al.
      Navigation-aided reconstruction of medial orbital wall and floor contour in cranio-maxillofacial reconstruction.
      ,
      • Cai E.Z.
      • Koh Y.P.
      • Hing E.C.H.
      • et al.
      Computer-assisted navigational surgery improves outcomes in orbital reconstructive surgery.
      ,
      • Schlittler F.
      • Schmidli A.
      • Wagner F.
      • et al.
      What is the incidence of implant malpositioning and revision surgery after orbital repair?.

      Clinics care points

      • Significant progress has been made in terms of reliability and accuracy of orbital reconstruction with the introduction of computer-assisted surgery techniques. A significant effect is seen for each individual technique and techniques can be combined to further optimize feedback during reconstruction.
      • The effect of intraoperative navigation can be enhanced by providing more intuitive feedback on implant position. The more intuitive and user friendly the feedback is presented, the more accurate the implant is positioned.
      • Translation of results from a preclinical setting to a clinical setting always warrants caution.
      • The focus in these studies, and many other studies, is on implant position. However, the main goal in orbital reconstruction is not to improve surgical precision but to improve quality of life for the patient. Accurate implant positioning should be viewed as a means to achieve the goal, rather than a goal in itself.

      Disclosure

      In the submission process, in kind funding was declared. In all these studies, Brainlab provided the navigation equipment in kind, while KLS Martin provided the preformed orbital implants in kind. Additionally, for the real-time navigation, Brainlab provided the IGT link license in kind. KLS Martin provided the instrument in kind. This sub-project was supported by the S.O.R.G. Research Grant Award 2017.

      References

        • Gellrich N.-C.
        • Schramm A.
        • Hammer B.
        • et al.
        Computer-assisted secondary reconstruction of unilateral posttraumatic orbital deformity.
        Plast Reconstr Surg. 2002; 110: 1417-1429
        • Bly R.A.
        • Chang S.-H.
        • Cudejkova M.
        • et al.
        Computer-guided orbital reconstruction to improve outcomes.
        JAMA Facial Plast Surg. 2013; 15: 113-120
        • Rana M.
        • Chui C.H.K.
        • Wagner M.
        • et al.
        Increasing the accuracy of orbital reconstruction with selective laser-melted patient-specific implants combined with intraoperative navigation.
        J Oral Maxillofac Surg. 2015; 73: 1113-1118
        • Zimmerer R.M.
        • Ellis E.
        • Aniceto G.S.
        • et al.
        A prospective multicenter study to compare the precision of posttraumatic internal orbital reconstruction with standard preformed and individualized orbital implants.
        J Craniomaxillofac Surg. 2016; 44: 1485-1497
        • Chen C.-T.
        • Pan C.-H.
        • Chen C.-H.
        • et al.
        Clinical outcomes for minimally invasive primary and secondary orbital reconstruction using an advanced synergistic combination of navigation and endoscopy.
        J Plast Reconstr Aesthet Surg. 2018; 71: 90-100
        • De Cuyper B.
        • Abeloos J.
        • Swennen G.
        • et al.
        Intraoperative navigation and cone beam computed tomography for restoring orbital dimensions: a single-center experience.
        Craniomaxillofac Trauma Reconstr. 2020; 13: 84-92
        • Dubois L.
        • Jansen J.
        • Schreurs R.
        • et al.
        Predictability in orbital reconstruction: a human cadaver study. Part I: endoscopic-assisted orbital reconstruction.
        J Craniomaxillofac Surg. 2015; 43https://doi.org/10.1016/j.jcms.2015.07.019
        • Jansen J.
        • Schreurs R.
        • Dubois L.
        • et al.
        The advantages of advanced computer-assisted diagnostics and three-dimensional preoperative planning on implant position in orbital reconstruction.
        J Craniomaxillofac Surg. 2018; https://doi.org/10.1016/j.jcms.2018.02.010
        • Jansen J.
        • Schreurs R.
        • Dubois L.
        • et al.
        Intraoperative imaging in orbital reconstruction: how does it affect the position of the implant?.
        Br J Oral Maxillofac Surg. 2020; 58: 801-806
        • Dubois L.
        • Schreurs R.
        • Jansen J.
        • et al.
        Predictability in orbital reconstruction: a human cadaver study. Part II: navigation-assisted orbital reconstruction.
        J Craniomaxillofac Surg. 2015; 43https://doi.org/10.1016/j.jcms.2015.07.020
        • Dubois L.
        • Essig H.
        • Schreurs R.
        • et al.
        Predictability in orbital reconstruction. A human cadaver study, part III: implant-oriented navigation for optimized reconstruction.
        J Craniomaxillofac Surg. 2015; 43: 2050-2056
        • Schreurs R.
        • Dubois L.
        • Ho J.
        • et al.
        Implant-oriented navigation in orbital reconstruction part II: preclinical cadaver study.
        Int J Oral Maxillofac Surg. 2020; 49: 678-685
        • Jaquiéry C.
        • Aeppli C.
        • Cornelius P.
        • et al.
        Reconstruction of orbital wall defects: critical review of 72 patients.
        Int J Oral Maxillofac Surg. 2007; 36: 193-199
        • Schreurs R.
        • Dubois L.
        • Becking A.G.
        • et al.
        Quantitative assessment of orbital implant position-a proof of concept.
        PLoS One. 2016; 11https://doi.org/10.1371/journal.pone.0150162
        • Eggers G.
        • Mühling J.
        • Marmulla R.
        Image-to-patient registration techniques in head surgery.
        Int J Oral Maxillofac Surg. 2006; 35: 1081-1095
        • Metzger M.C.
        • Hohlweg-Majert B.
        • Schön R.
        • et al.
        Verification of clinical precision after computer-aided reconstruction in craniomaxillofacial surgery.
        Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2007; 104: e1-e10
        • Widmann G.
        • Stoffner R.
        • Bale R.
        Errors and error management in image-guided craniomaxillofacial surgery.
        Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009; 107: 701-715
        • Schramm A.
        • Suarez-Cunqueiro M.M.
        • Rücker M.
        • et al.
        Computer-assisted therapy in orbital and mid-facial reconstructions.
        Int J Med Robot Comput Assist Surg. 2009; 5: 111-124
        • Yu H.
        • Shen S.G.
        • Wang X.
        • et al.
        The indication and application of computer-assisted navigation in oral and maxillofacial surgery—Shanghai’s experience based on 104 cases.
        J Craniomaxillofac Surg. 2013; 41: 770-774
        • Gander T.
        • Essig H.
        • Metzler P.
        • et al.
        Patient specific implants (PSI) in reconstruction of orbital floor and wall fractures.
        J Craniomaxillofac Surg. 2015; 43: 126-130
        • Bevans S.E.
        • Moe K.S.
        Advances in the reconstruction of orbital fractures.
        Facial Plast Surg Clin. 2017; 25: 513-535
        • Moe K.S.
        • Murr A.H.
        • Wester S.T.
        Orbital fractures.
        Facial Plast Surg Clin. 2018; 26: 237-251
        • Schmelzeisen R.
        • Gellrich N.C.
        • Schoen R.
        • et al.
        Navigation-aided reconstruction of medial orbital wall and floor contour in cranio-maxillofacial reconstruction.
        Injury. 2004; 35: 955-962
        • Mahoney N.R.
        • Peng M.Y.
        • Merbs S.L.
        • et al.
        Virtual fitting, selection, and cutting of preformed anatomic orbital implants.
        Ophthal Plast Reconstr Surg. 2017; 33: 196-201
        • Stuck B.A.
        • Hülse R.
        • Barth T.J.
        Intraoperative cone beam computed tomography in the management of facial fractures.
        Int J Oral Maxillofac Surg. 2012; 41: 1171-1175
        • Wilde F.
        • Schramm A.
        Intraoperative imaging in orbital and midface reconstruction.
        Facial Plast Surg. 2014; 30: 545-553
        • Schreurs R.
        • Dubois L.
        • Becking A.G.
        • et al.
        Implant-oriented navigation in orbital reconstruction. Part 1: technique and accuracy study.
        Int J Oral Maxillofac Surg. 2018; 47https://doi.org/10.1016/j.ijom.2017.09.009
        • Essig H.
        • Dressel L.
        • Rana M.
        • et al.
        Precision of posttraumatic primary orbital reconstruction using individually bent titanium mesh with and without navigation: a retrospective study.
        Head Face Med. 2013; 9: 1-7
        • Rana M.
        • Holtmann H.
        • Kanatas A.N.
        • et al.
        Primary orbital reconstruction with selective laser melted core patient-specific implants: overview of 100 patients.
        Br J Oral Maxillofac Surg. 2019; 57: 782-787
        • Zimmerer R.M.
        • Gellrich N.-C.
        • von Bülow S.
        • et al.
        Is there more to the clinical outcome in posttraumatic reconstruction of the inferior and medial orbital walls than accuracy of implant placement and implant surface contouring? A prospective multicenter study to identify predictors of clinical outcome.
        J Craniomaxillofac Surg. 2018; 46: 578-587
        • Jansen J.
        • Dubois L.
        • Maal T.J.J.
        • et al.
        A nonsurgical approach with repeated orthoptic evaluation is justified for most blow-out fractures.
        J Craniomaxillofac Surg. 2020;
        • Cai E.Z.
        • Koh Y.P.
        • Hing E.C.H.
        • et al.
        Computer-assisted navigational surgery improves outcomes in orbital reconstructive surgery.
        J Craniofac Surg. 2012; 23: 1567-1573
        • Schlittler F.
        • Schmidli A.
        • Wagner F.
        • et al.
        What is the incidence of implant malpositioning and revision surgery after orbital repair?.
        J Oral Maxillofac Surg. 2018; 76: 146-153